BATH
|

|
|
|
Bath Circulator
|
|
|
|
|

External and internal circulating with
powerful pumping capacity CW-G series is designed for controlling temperature in external instrumentation, or for immersion application. Pump pressure is adjustable in five stages and all models come with built-in tab water cooling coil.
Performance Features
¢º Operating features • 5¡É above room temperature to 150¡É • Microprocess PID control/ Auto-tuning/ Calibration • Digital timer: 1 min to 99 hr 59 min, delayed ON/ OFF • Over-temperature limiter and low fluid level protection • 3 different temperature values memorable without auto-tuning • Digital LED display (0.1¡É resolution) with splash-proof keypads • RS-232 interface available
¢º Constructional features • Exclusive variable-speed pump to liquid viscosity for accurate, stable, and efficient temperature management (1 to 3 steps of pump for internal circulation/ 4 and 5 steps for external circulation) • Powerful pressure and suction pumps provide contrast pressure and flow rate over the full
temperature range • Baffle plate screens circulating pump, heater and sensor to minimize shaking bath fluids and protect sample
and operators from direct contact • Dry-running protection; automated heating element cut-off and audible and visual warning • Seamless stainless steel tank with a reservoir drain that resists corrosion and make cleaning easy • Space saving design uses minimal bench space
Specifications
Model |
CW-05G |
CW-10G |
CW-20G |
CW-30G |
Bath volume (L/cu ft) |
5.0 / 0.2 |
10.0 / 0.4 |
20.0 / 0.7 |
30.0 / 1.1 |
Working Temperature Range2)(¡É/¢µ) |
Amb. +5 to 150 / Amb. +9 to 302 |
Temperature Stability1)(¡¾¡É/¢µ) |
0.05 / 0.09 |
0.05 / 0.09 |
0.1 / 0.18 |
0.1 / 0.18 |
Pump |
Max. pressure, max. flow rate |
5.9 PSI (406.8 millibar), 7.4 gallons/minute (28.0 liters/minute) |
Max. suction flow rate |
2.6 gallons/minute (10.0 liters/minute) |
Dimensions |
Bath opening/depth (W¡¿L,D) (mm/inch) |
150¡¿99, 160 / 5.9¡¿3.9, 6.3 |
211¡¿154, 180 / 8.3¡¿6.1, 7.1 |
290¡¿214, 200 / 11.4¡¿8.4, 7.9 |
300¡¿264, 230 / 11.8¡¿10.4, 9.0 |
Overall (W¡¿L¡¿H) (mm/inch) |
220¡¿395¡¿424 / 8.7¡¿15.6¡¿16.7 |
281¡¿450¡¿444 / 11.1¡¿17.7¡¿17.5 |
360¡¿510¡¿464 / 14.2¡¿20.1¡¿18.3 |
370¡¿560¡¿494 / 14.6¡¿22.0¡¿19.4 |
Net weigth (kg/lbs) |
12.5 / 27.6 |
15.0 / 33.0 |
17.5 / 38.6 |
19.5 / 43.0 |
Electrical Requirements (230V, 60Hz) |
50/60 Hz, 4.5A |
50/60 Hz, 6.7A |
50/60 Hz, 8.8A |
50/60 Hz, 8.8A |
Electrical Requirements (120V, 50Hz) |
50/60 Hz, 8.6A |
50/60 Hz, 8.6A |
50/60 Hz, 8.6A |
50/60 Hz, 8.6A | 1) If you operate the unit in 1 to 3 steps of pump, the temperature range can be from Amb. +5¡É but in 4 and 5 steps, the range will be from Amb. +15¡É. 2) The heat-up time is measured on a 230V electrical condition with 1000W heater power.
Accessories & Options
 |
 Highly durable stainless steel racks with adjustable angle and tube height support. Do not need any other base plates to use it, simply release and mount racks directly. | |
 |
 Ideal for holding a wide variety of tubes, flasks, and other glassware. All component are made of stainless steel. (for all models) | |
 |
 Additional suction function for powerful external/ internal circulating. Maximum suction flow rate is 10.0 liters per minute. (only for CW-G and RW-G series) * The additional suction function is needed to have a prior request. | | |
|
|
|
|
|
|
|
|
|
|
|
|

|
|

|